Configuration
80
Ventilation Defaults
Apnoea ventilation on/off
To determine whether apnoea ventilation is automatically ready
for use when starting.
● Press menu button »Calib./Config.«.
● Press menu button »Ventilation jj«.
● Enter code number »3032«.
● Select the menu »Mode« via the menu button
»Ventilation
jj«.
● Select screen key »Apnoea vent.« = turn rotary knob.
● Switch on apnoea ventilation = press rotary knob,
the "LED" in the screen key »Apnoea vent.« lights up yellow.
● Switch off apnoea ventilation = press rotary knob,
the "LED" in the screen key »Apnoea vent.« lights up green.
Leakage compensation* on/off
Automatic leakage compensation allows the unit to
compensate leakages up to 100 % of the set tidal volume in all
volume-controlled ventilation modes.
The setting for »Leakage compensation on/off« is saved and
remains effective when the unit is restarted.
● Press menu button »Calib./Config.«,
● Press menu button »Ventilation jj«.
● Enter the numerical code »3032«.
● Select the menu »Parameter« via the menu button
»Ventilation
jj«.
Display:
● Select screen key »Leakage comp.«,
● Switch on »Leakage comp.« = press rotary knob,
the "LED" in the screen key »Leakage comp.« lights up
yellow.
● Switch off »Leakage comp.« = press rotary knob,
the "LED" in the screen key »Leakage comp.« lights up
green.
* For a detailed description of leakage compensation, refer to
page 124 in the Annex.