91
Care
Removing parts
Medicament nebuliser (option)
1 Disconnect nebuliser hose from medicament nebulizer and
from the port on the unit.
2 Disconnect the medicament nebuliser from the adult hose
system
or
2 from the paediatric hose system.
3 Remove catheter connector (ISO cone Ø
15 / Ø 11) from
the inlet.
4 Remove the adapter (ISO cone Ø
22 / Ø 11) from the outlet.
5 Remove the corrugated hose from the adapter.
Dismantle the medicament nebuliser in accordance with its
Instructions for Use.
Prepare the individual parts of the medicament nebuliser
and the adapter parts for cleaning and disinfection in a
cleaning and disinfection machine.
Ventilation hoses
Remove ventilation hoses from the device ports.
Remove the water traps from the ventilation hoses.
Remove the water containers from the water traps.
Prepare the ventilation hoses, water traps and their
collecting jars and the Y-piece for cleaning and disinfection
in a cleaning and disinfection machine.
Flow-Sensor
6 Push the flow sensor to the left as far as it will go and
7 pull out.
The flow sensor cannot be disinfected/cleaned in a
machine, nor can it be sterilised by the hot steam method.
See page 95 for cleaning instructions of the flow sensor.
Disinfect flow sensor for about 1 hour in 70 % ethanol
solution.
The flow sensor may be re-used as long as calibration can
be carried out successfully.
Expose the sensor to air for at least 30 minutes to allow
the alcohol to evaporate.
Otherwise the sensor could be damaged beyond control
by ignition of any residual alcohol during calibration.
042
1
2
043
5
4
2
1
3
044
7
6