Care
98
Maintenance Intervals
Clean or replace cooling air filter
Filter must be cleaned or replaced when soiled or at the
latest after 4 weeks.
Replace after 1 year at the latest.
1 Remove cooling-air filter from its slot on the back of
machine.
Replace or clean in warm water with added detergent;
dry well.
Insert cooling-air filter in slot, taking care not to crease it.
Dispose of used cooling-air filter with domestic waste.
Note the Instructions for Use of the option when using the
(optional) DC power supply MB.
Removing and reinserting ambient-air filter
Filter must be cleaned or replaced when soiled or at the
latest after 4 weeks.
Replace filter every year.
2 Swivel port to the left.
3 Unscrew the screw (e.g. with a coin), and remove the
protective cover.
4 Remove the ambient-air filter from the protective cover.
Push the new/cleaned ambient-air filter under the lugs.
Replace protective cover, and tighten screw with a coin.
Dispose of used ambient-air filter with domestic waste.