5664.500/800 Evita 4/Evita 2 dura 06/98 Repair Instructions Page 31
For internal use only. Copyright reserved.
GBR5664500T055664800.fm 18.03.99
Dräger Medizintechnik
D
4 Paediatric Flow PCB Option for Evita 4 as of SW 2.20
4.1 Power Supply to the Paediatric Flow PCB
4.2 Connector Pin Assignment of the Paediatric Flow PCB
The connectors X1 and X2 are not considered in this connector pin assignment.
X2 (Power supply only)
Pin Assignment Pin Assignment
1c +12 V 8c AGND
2c +24 V 15c DGND
3c +15 V 16c +5 V
4c –15 V
X3 (Flow sensor)
Pin Assignment Pin Assignment
1 Z-Flow-01 4 Z-Flow-07
2Z-Flow-038 Housing
3 Z-Flow-05 15 Housing
X4 (Analog voltage test)
Pin Assignment Pin Assignment
1 +15 V-A 4 –15 V-A
2 +5/15 V-B 5 AGND
3 +10 V-A
CAUTION
Electrostatic discharge can damage electronic components. Use an ESD
protection mat and a wrist strap when handling electronic modules.