5664.500/800 Evita 4/Evita 2 dura 06/98 Repair Instructions Page 45
For internal use only. Copyright reserved.
GBR5664500T055664800.fm 18.03.99
Dräger Medizintechnik
D
6.5 Replacing the Power Supply PCB
Switch the Evita off.
Remove the CO
2
Carrier PCB.
Remove the screws which secure the Power Supply PCB and remove the
Power Supply PCB.
Mount the new Power Supply PCB on the CO
2
Carrier PCB.
Install the CO
2
Carrier PCB in the Evita.
Secure the CO
2
Carrier PCB with the screws.
Check the Evita by following the instructions in the Evita 4/Evita 2 dura Test Certificate.