5664.500/800 Evita 4/Evita 2 dura 06/98 Repair Instructions Page 53
For internal use only. Copyright reserved.
GBR5664500T055664800.fm 18.03.99
Dräger Medizintechnik
D
X4 (Display signal)
Pin Assignment Pin Assignment
1 GND 25 GND
22 HSYNCH (horizontal
synchronization)
26 GND
23 GND 28 +5 V
24 VSYNCH (vertical synchronization) 32 +12 V
X7 (Display power)
Pin Assignment Pin Assignment
1 +12 V 5 DIMPWM (signal to dim the
inverter)
2 GND 7 +5 V
3+24 V
X10 (Background debug mode interface)
Pin Assignment Pin Assignment
1 GND 5 RESETQ (reset signal)
2 BKPT (break input) 6 IFETCH (status signal from 68332)
3 GND 7 +5 V
4 FREEZE (status signal from
68332)
8 IPIPE (status signal from 68332)
X14 (Membrane keyboard)
Pin Assignment Pin Assignment
23 GND 27 BUTTON (button of the control
knob)
24 +5 V 28 ALARMLED (anode of the alarm
LEDs)
25 KN1 (signal 1 from the control
knob)
31 LS1 (speaker connection +)