15.2.3 X5 Connector on the Frontpanel PCB to the Control Knob
15.2.4 X7 Connector on the Frontpanel PCB to the display
15.2.5 X9 Connector on the Frontpanel PCB to the DC/AC Transformer
15.3 Front Panel PCB Repair Information
• Decoder D48 (only present in colour display option) should be taken from the old
Frontpanel PCB and mounted on the new Frontpanel PCB.
• After replacing a printed circuit board, download the respective software into the
Frontpanel PCB.
Pin Assignment
1/2 GND
6 +5 V
Pin Assignment
8/10/12/14/16/1
8/20/22
GND
24/26 UBKL +13 V (display operating voltage, delayed when switching on
the Evita 2 dura)
28/30 +5 V
32 –UK –32 V to –35 V, depending on contrast setting of the display
34 +UK +32 V to +35 V, depending on contrast setting of the display
Pin Assignment
1+24 V
4 +5 V
5UBKL