5664.500/800 Evita 4/Evita 2 dura 06/98 Repair Instructions Page 77
For internal use only. Copyright reserved.
GBR5664500T055664800.fm 18.03.99
Dräger Medizintechnik
D
X3 (Test points)
Pin Assignment Pin Assignment
1 +5 V-A 5 +15 V-A
2 –5 V-A 6 +10 V-A
3–15 V-A7 AGND
4+15 V-C
X4 (Flow sensor)
Pin Assignment Pin Assignment
1 FGND 5 H-DR-OUT2
2 R-121-OUT 6 U-SENSOR
3 K-DR-OUT 7 H-DR-OUT1
4 R-15-OUT
X5 (PEEP/PIP valve)
Pin Assignment Pin Assignment
1 5 V-B 2 PEEP-OUT
X6 (O
2
sensor)
Pin Assignment Pin Assignment
1 +15 V-B 3 O2-IN
2 AGND 4 –15 V-A
X7 (Inspiration/expiration pressure sensor)
Pin Assignment Pin Assignment
1AGND3 PEXP
2 PINSP 4 +15 V-F