5664.500/800 Evita 4/Evita 2 dura 06/98 Repair Instructions Page 85
For internal use only. Copyright reserved.
GBR5664500T055664800.fm 18.03.99
Dräger Medizintechnik
D
Mount the new Pneumatics Controller PCB by following the reverse method used for
dismounting.
Load the relevant software into the Pneumatics Controller PCB.
Mount the rear panel.
Calibrate the PEEP valve (see “Calibrating the PEEP Valve”).
Check the Evita by following the instructions in the Evita 4/Evita 2 dura Test Certificate.
16.4 Tests possible in DS mode
16.4.1 Diagnosis mode
refer to section of "Repair Instructions - Service No./ Modi, Error list" chapter 3.4,
“Pneumatic” diagnosis, page 74.
16.4.2 External DS mode using PC
refer to section of "Repair Instructions - Service No./ Modi, Error list" chapter 9.3.1, Valve
switching, page 128.
refer to section of "Repair Instructions - Service No./ Modi, Error list" chapter 9.3.2, Mixer,
page 129.
refer to section of "Repair Instructions - Service No./ Modi, Error list" chapter 9.3.3,
Setpoint input for PEEP valve, page 129.
refer to section of "Repair Instructions - Service No./ Modi, Error list" chapter 9.3.4, Sen-
sors, page 130.
refer to section of "Repair Instructions - Service No./ Modi, Error list" chapter 9.3.5,
Voltages, page 134.
refer to section of "Repair Instructions - Service No./ Modi, Error list" chapter 9.3.6, Cali-
bration, page 136.
refer to section of "Repair Instructions - Service No./ Modi, Error list" chapter 9.3.7, CPU,
page 137.