If Capno Plus option is installed:
Checking / calibrating CO
2 sensor
The CO2 sensor is works-calibrated and can be used
without further calibration on any Evita 4 unit.
A CO2 zero calibration is performed as part of the device
check.
Calibration of the CO2 sensor is only required if:
on checking the calibration with a test filter or with
test gas, the specified test values are not met
and
on half-yearly device inspections.
The calibration check or calibration can be performed
during ventilation.
CO2 zero calibration
if the device requests CO2 zero calibration with the
screen message: CO2 zero calibration?
if the CO2 curve no longer returns to zero after each
inspiration
before each calibration test, page 102 onwards
before each CO2 calibration, page 105.
Switch on Evita 4. Wait about 3 minutes for
completion of the warm-up phase of the CO2 sensor.
After about 3 minutes, the measured values will be within
the specified accuracy.
Press the »Calibration« key.
Display (example):
Touch the »CO2« screen key.
Touch the »Zero« screen key.
Message:
Park CO2 sensor
Confirm = press rotary knob.