119
Configuration
Ventilation
Start-up values for ventilation parameters
and alarm limits
– To set the ventilation parameters and alarm limits you
would like to be activated on switching on the device.
Setting start-up values for ventilation
parameters »V
T, f«
The start-up values for the tidal volume (VT) and
frequency (f) required for the patient are determined by
Evita 4:
either as a function of the ideal body weight
or
as a function of the patient mode
(paediatrics or adults).
● Press »Configuration« key.
● Touch the »Ventilation« screen key.
Enter access code 3032:
● Touch the corresponding screen keys.
● Touch the »Start up settings« screen key.
Display (example):
To determine the start-up values of the ventilation
parameters VT, f on the basis of the ideal body weight:
● Touch the screen key »VT, f by weight« and confirm
with the rotary knob.
To determine the start-up values of the ventilation
parameters VT, f on the basis of the patient mode:
● Touch the screen key »VT, f by patient« and confirm
with the rotary knob.
179 37 207