Performance Data
Control principle time-cycled, volume-constant, pressure-controlled
Intermittent PEEP frequency 2 cycles every 3 minutes
Medicament nebulisation for 30 minutes
Bronchial suction
disconnection detection automatic
reconnection detection automatic
oxygen enrichment max. 3 minutes
active suction phase max. 2 minutes
final oxygen enrichment 2 minutes
Valve response time T0...90 ≤5 ms
Supply system for spontaneous breathing
and ASB adaptive CPAP system with high initial flow
max. flow rate 2 L/s in 8 ms
max. inspiratory flow 180 L/min
Equipment compliance
(with humidifier Aquapor and
patient tubing system for adults) ≤2 mL/mbar
Insp. Resistance ≤2.3 mbar/L/s
Exp. Resistance ≤3.8 mbar/L/s
Dead Space Volume incl. CO2-cuvette ≤16 mL
Equipment compliance
(with Fisher & Paykel MR 730 humidifier
and tubing system paediatric) ≤1 mL/mbar
Insp. Resistance ≤4.1 mbar/L/s
Exp. Resistance ≤4.1 mbar/L/s
Dead Space Volume incl. CO2-cuvette ≤6 mL
Additional functions
Inspiratory relief valve opens if medical air supply fails (pressure <1.2 bar),
enables spontaneous breathing with filtered ambient air.
Safety valve opens the breathing system at 100 mbar.
Measured value displays
Airway pressure measurement
Max. airway pressure Ppeak
Plateau pressure Pplat
Pos. end-exp. pressure PEEP
Mean airway pressure Pmean
Min. airway pressure Pmin
Range 0 to 99 mbar
Resolution 1 mbar
Accuracy ±2 mbar
O2 measurement in main flow (inspiratory side)
Inspiratory O2 concentration FiO2
Range 15 to 100 Vol.%
Resolution 1 Vol.%
Accuracy ±3 Vol.%
153
Technical Data