20
Screen Knobs for Setting Parameters
Display (example):
e.g. »PEEP« screen knob.
1 Touch the »PEEP« screen knob:
It changes colour from green to yellow =
setting function selected.
2 Turn rotary knob = Adjust setting. The value is
displayed in the knob.
3 Press rotary knob = Confirm. The knob changes
colour from yellow to green, and the setting is
validated and takes effect.
While pressure values, such as Pmax, are being set, they
are displayed in the Paw (t) curve as a dashed black line.
To cancel the setting:
● Press the screen knob again
or
● press another screen knob.
1
2
3
FiO
2
21
P
peak
33 >
50
P
plat
27
MV 5.3 _
7.0
4.0
IPPV
BIPAP
SIMV
ASB
Extra
settings
Basic
settings
IPPV
Assist
A
m
21
O2 [%]
.500
VT
1.7
Tinsp
21
f
60
Flow
5
PEEP
60
40
20
0
-10
0 2 4 6 8 t [s]
P
aw
[mbar]
Operating Concept
On-Screen Controls
PEEP
5
10
PEEP [mbar]
PEEP [mbar]
1
3
2