The following tests are performed during the device
check:
Correct operation of auxiliary and power failure alarms
Seating and clear passage of the expiratory valve
Seating of the flow sensor
Seating of the neonate flow sensor
(if "NeoFlow" option is installed)
Type of humidifier
Completeness of hose system
Function of the air-O2 changeover valve
Function of the safety valve
Gas supply
Calibration of the flow sensor
Calibration of the neonate flow sensor
(if "NeoFlow" option is installed)
Calibration of the O2 sensor
Leakproofing of the hose system
On completion of the device check, a checklist is
displayed on the screen to show the results of the check.
Correct result :
Incorrect result : F
Check not performed : – –
In the event of incorrect results, e.g. if the hose system is
not sufficiently leakproof:
Eliminate the cause of the fault
Touch the »Repeat check« screen key
Only the tests with incorrect results are repeated.
After successful completion of the device check,
Evita 4 is ready for operation.
Either:
immediately start up Evita 4 by pressing key »O «
or:
leave Evita in standby mode
or:
switch off Evita for later use.
Switch on back panel = pivot flap to the side and
press button in fully and release.