The screen key »prev. patient« can be used to restore
the specific patient settings, including alarm limits and
monitoring status, effective before switching off the
device.
Example:
Previous patient
The previous modes are displayed in the status line:
– Previous ventilation mode
– Previous patient mode
– Previous application mode (tube or mask for
optional NIV)
The key »prev. patient« is not displayed by Evita 4
following a loss of data or removal of a previously used
option (e.g. NeoFlow), thus preventing restoration of the
previous setting.
Restoration of the previous setting is similarly prevented
by Evita 4 if it was configured in such a way before
switching off that the former patient mode is no longer
available.
Selecting the patient mode
if configured
EITHER
● the »Adults« key or
the »Paed.« key or
the »Neo.« key (NeoFlow option) and
enter the ideal body weight
if configured
With the ideal body weight, Evita 4 determines the
start-up settings of the ventilation parameters. The
start-up value is marked on the relevant screen knob
by an arrow ( j).
● Touch the screen knob.
● Enter the ideal body weight [kg] with the manual
dial-knob = turn rotary knob.
● Confirm the setting = press rotary knob.
OR
Select the previous settings
● Touch the key »prev. patient«.
● Confirm = press rotary knob.