53
Operation
Setting Ventilation Modes
Setting Ventilation Modes
The ventilation modes IPPV, BIPAP, SIMV and ASB are
already configured in the unit. If other ventilation modes
are used, please refer to page 113 "Selecting ventilation
modes".
IPPV
Intermittent Positive Pressure Ventilation
Volume-controlled ventilation with fixed, mandatory
minute volume MV and user-adjusted tidal volume VT
and frequency f.
For patients having no spontaneous breathing.
Set the pattern of ventilation for IPPV with the ventilation
parameters:
Tidal volume »VT«
Insp. Flow »Flow«
Frequency »f«
Inspiration time »Tinsp«
O2 concentration »O2«
Positive end-expiratory pressure »PEEP«
––––––––––––––––––––––––––––––––––––––––––––––––
To set:
● Touch the relevant screen knob.
● Adjust value = turn rotary knob.
● Confirm setting = push rotary knob.
––––––––––––––––––––––––––––––––––––––––––––––––
Paw
Flow
T
insp Te
Pplat
Insp.
pause
Pmax
(configu-
rable)
PEEP
1
f
Insp. flow
t
t
065 37 207