EasyManua.ls Logo

Dräger Evita 4 - Page 61

Dräger Evita 4
192 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BIPAP and BIPAP/ASB can be expanded with the
following ventilation parameters:
Flowtrigger – for synchronisation with attempted
spontaneous breathing by the patient.
By setting on the flow trigger level, the mandatory strokes
are synchronised with the patient's spontaneous
breathing attempts.
Spontaneous breathing activity by the patient is indicated
by the brief display of a lung symbol instead of the usual
symbol for the patient mode.
Apnoea Ventilation – for automatic switch-over to
volume-controlled mandatory ventilation if the patient
stops breathing.
If breathing stops, Evita 4 emits an alarm after the set
alarm time (TApnoea > ) and starts volume-controlled
ventilation with the set ventilation parameters:
Frequency »fApnoea«
Tidal volume »VTApnoea«
––––––––––––––––––––––––––––––––––––––––––––––––
To set (Example: Flowtrigger)
Touch the »Extra settings« screen key.
Touch the screen key corresponding to
»Flowtrigger«.
Set values = touch the »FlowTrig« screen knob, and
turn and press the rotary knob.
––––––––––––––––––––––––––––––––––––––––––––––––
61
Operation
Setting Ventilation Modes
079 37 207

Table of Contents

Other manuals for Dräger Evita 4

Related product manuals