The medicament nebuliser is automatically switched off
after 30 minutes.
After administration of the aerosol, the flow sensor is
automatically cleaned and calibrated in order to prevent
malfunctions in flow measurement.
Only use medicament nebuliser 84 12 935
(white central body).
Prepare the medicament nebuliser in accordance with its
Instructions for Use.
If other pneumatic medicament nebulisers are used,
there may be major deviations in tidal volume and
inspiratory O2 concentration.
For use during adult ventilation
1 Connect the nebuliser to the inspiratory side
(temperature sensor side) of the Y-piece.
2 Connect the inspiration hose to the medicament
nebuliser.
● Place the medicament nebuliser in the vertical
position.
● Using clamps, route the nebuliser hose back to the
ventilator along the expiratory hose.
For use during paediatric ventilation
3 Insert the catheter connector (ISO cone Ø15 / Ø11)
in the inlet of the medicament nebuliser.
4 Insert the adapter (ISO cone Ø22 / Ø11) in the outlet.
5 Fit the corrugated hose (0.13 m long) to the outlet
adapter.
6 Remove the corrugated hose of the hose set from the
inspiratory adapter of the Y-piece and connect it to
the inlet adapter of the medicament nebuliser.
7 Connect the free end of the corrugated hose at the
outlet of the medicament nebuliser to the inspiratory
adapter of the Y-piece.
89
1
2
3
4
5
6
7
Operation
Special Functions
Medicament nebulisation