Selecting Standby Mode
No ventilation takes place in standby mode!
Uses of Standby:
– to perform the device check
– to maintain Evita 4 ready for operation while the
patient is absent.
– to change patient mode.
1 Hold down the »O« key for about 3 seconds.
An alarm tone sounds after switching on standby.
To switch off the standby alarm tone:
2 Press »Alarm Reset« key.
The standby alarm tone cannot be cancelled with
the »g« key.
Display (example):
If the patient mode or ideal body weight should be
changed during Standby, Evita 4 will determine new start
values for ventilation, see page 50.
Terminating standby mode
– to continue ventilation.
1 Press »O« key.
The LED goes out, and the current ventilation
parameters are again effective.
Operation
Selecting Standby Mode