Getting started
Display of results of the breathing circuit check
Prerequisite: The System check page (A) is
opened.
z Touch the Check results tab (B).
The detailed results of the check are displayed.
Checking the switch to battery operation
z Unplug the mains plug.
The device switches to battery operation without in-
terruption. The Battery activated alarm is dis-
played.
z Plug the power cable back in.
The device switches back to mains operation. The
Battery activated alarm message goes out.
Checking the alarm signaling
When the system check has been successfully com-
pleted, the device is ready for operation. The alarm
signaling can also be checked. The description of
alarm signaling can be found in the instructions for
use in chapter "Alarms." Additional information on
alarm criteria can be found in the instructions for use
in chapter "Alarm – Cause – Remedy."
High-priority alarm message
1 Start ventilation.
2 After 2 minutes set the upper alarm limit for
MVe to a value below the measured value MVe.
The MV high alarm is triggered.
Medium-priority alarm message
For Evita Infinity V500 and Evita V300:
1 Start ventilation.
2 Set the upper alarm limit for VT to a value below
the measured value VT.
The VT high alarm is triggered.
For Babylog VN500:
1 Start ventilation.
2 Switch on the additional setting Volume Guar-
antee
.
3 Reduce the upper alarm limit for Paw until the
VT not reached, Pmax active alarm is trig-
gered.
Low-priority alarm message
1 Start ventilation.
2 In the Special maneuvers > Maneuvers dialog
window, press and hold the Man. insp./hold
button until the Inspiratory hold interrupted
alarm is triggered.
103_2
C Compliance [mL/mbar]
D Flow [L/min]
E Inspiratory resistance [mbar/L/s]
F Expiratory resistance [mbar/L/s]
Start/Standby
C
A
B
DEF