Operation
Medication nebulization
Applies to:
Evita Infinity V500
Babylog VN500
Evita V300
Safety information on medication
nebulization
Continuous medication nebulization
Prerequisite: The Special maneuvers >
Nebulization page is open.
z Press the (A) button.
Continuous medication nebulization is started. The
message Continuous nebulization in progress.
is displayed.
Medication nebulization is interrupted every
30 minutes and the flow sensor is calibrated. After
the flow sensor has been calibrated, medication
nebulization is continued.
When medication nebulization is used in the patient
category Neo. or Ped. pat. and the neonatal flow
sensor has thus been removed, medication nebuli-
zation is not interrupted.
If the parameter field for continuous nebulization
Cont. neb. has been configured for display, the du-
ration of medication nebulization is displayed.
Mains power supply / DC power supply
Applies to:
Evita Infinity V500
Babylog VN500
Evita V300
Battery charging
The battery charge display applies to both charging
and discharging.
CAUTION
Ventilation impaired
If unapproved pneumatic medication nebulizers
are used, the actually delivered tidal volume and
O2 concentration may deviate from the displayed
values.
Only use the medication nebulizers listed in the
current list of accessories.
CAUTION
Ventilation impaired
Aerosols can impair the proper functioning of the
expiratory valve.
When using medication nebulization, shorten the
reprocessing cycles for the expiratory valve.
086_neu