Cleaning, disinfection and sterilization
Cleaning, disinfection and sterilization
Applies to:
Evita Infinity V500
Babylog VN500
Evita V300
Safety information on reprocessing
The components filled with contaminated gas dur-
ing normal operation and in the event of a fault must
be reprocessed.
During normal operation the expiratory valve, along
with the ejector and muffler, are filled with contami-
nated gas.
In the event of a fault, the inspiratory unit may be
contaminated.
Additional information on reprocessing can be
found in the instructions for use.
Reprocessing list
Semi-critical medical devices
Safety information concerning the reprocessing of
the neonatal flow sensor:
Items which can
be reprocessed
Recom-
mended re-
processing in-
tervals
Pre-
cleaning
Machine
cleaning and
disinfection
Manual Steriliza-
tion
Cleaning Disinfec-
tion
Breathing hoses Per patient/
weekly According to the corresponding instructions for use
NOTE
Do not use brushes for reprocessing the
sensor insert and do not use a syringe on the
sensor insert.
For reprocessing the housing use lint-free
brushes only.