Fabius
®
Repair Instructions
3-94
Copyright 2000 Draeger Medical, Inc.
sec3_repairinstruc_c.fm
16.5 APL Valve
In case of repair, replace the complete APL valve.
16.5.1 APL Valve Components
Fig. 79: APL valve components
Key
16.5.2 Replacing the APL Valve
• Remove the APL valve ring nut from the old APL valve.
• Mount the APL valve ring nut to the new APL valve.
• Mount the new APL valve to the Cosy compact breathing system mount.
• Perform a function check on Fabius.
1 APL valve ring nut
2 Cosy APL valve
3 Crater screw
3
2
1
RETURN TO CD-ROM TABLE OF CONTENTS