Fabius
®
Repair Instructions
3-31
Copyright 2000 Draeger Medical, Inc.
sec3_repairinstruc_c.fm
14.5 Power Transformer
In case of repair, replace the complete power transformer.
14.5.1 Testing the Power Transformer Output Voltage
• Switch control box off using ON/OFF switch (ON/OFF switch is located on rear
panel of control box).
• Unplug power cord from AC outlet.
• Check mains fuses of the control box (mains fuses are located on rear panel of
control box).
• If mains fuses are OK, proceed as follows:
• Remove control box from trolley (see control box chapter "Removing the Control
Box from the Gas Box").
• Remove lateral screws 1 from control box 2 and lift top cover 3 in direction of A
and away from control box 2.
• Place top cover 3 of control box 2 aside.
Fig. 13: Removing top cover of the control box
Electrostatic discharge can damage electrostatic sensitive
devices. Use a static-dissipative mat and a wrist strap when
handling electrostatic sensitive devices.
1..2..
Menu
PEEPIPPV
man./
spont.
Pmax T
I:TE TIP:TIfIPPVVT
IPPV
Paw [mbar]
FiO2
MV
Paw
60
40
20
0
24
7.9
28
0
80
21
12.0
3.0
40
8
1
1
A
2
3
RETURN TO CD-ROM TABLE OF CONTENTS