14.5 Power Transformer
In case of repair, replace the complete power transformer.
14.5.1 Testing the Power Transformer Output Voltage
Switch control box off using ON/OFF switch (ON/OFF switch is located on rear
panel of control box).
Unplug power cord from AC outlet.
Check mains fuses of the control box (mains fuses are located on rear panel of
control box).
If mains fuses are OK, proceed as follows:
Remove control box from trolley (see control box chapter "Removing the Control
Box from the Gas Box").
Remove lateral screws 1 from control box 2 and lift top cover 3 in direction of A
and away from control box 2.
Place top cover 3 of control box 2 aside.
Fig. 13: Removing top cover of the control box
Electrostatic discharge can damage electrostatic sensitive
devices. Use a static-dissipative mat and a wrist strap when
handling electrostatic sensitive devices.