28-26 DELTA/DELTA XL/KAPPA VF8
28 P
ERIPHERAL DEVICES AND ASSOCIATED SOFTWARE
The Incub2 menu has the following selections.
Menu Item Description Available Settings
Show All
Parameters
Displays all parameters
# Parameter
Boxes
Select number of Parameter
Boxes
1, 2 (default)
NOTE: Incub2 Parameter box is not
displayed when the number of Parameter
Boxes is set to 1
Parameter 1 Selects up to 3 parameters for
display in Incubator parameter box
T1skin, T2skin (default), Tair, %O2,
R.H., Wt, None
Parameter 2 T1skin, T2skin, Tair, %O2 (default),
R.H., Wt, None
Parameter 3 T1skin, T2skin, Tair, %O2, R.H., Wt
(default), None
Incubator
Mini Trend
Allows you to select parameters
for display in mini trends (See
page 6-7 for more information on
mini trend display).
NOTE: Both Incub1 & Incub2 can
display a mini trend.
T1skin, T2skin, Tair, %O2 (default),
R.H., Wt
Wt This is a read-only field. Only the
time, month, and day appears with
the Weight label
Not applicable