B-20 DELTA/DELTA XL/KAPPA VF8
B T
ECHNICAL DATA
8-Channel Analog Output
Number of channels: 8
Channel assignment: Follows ISD display order
Channel offset: <±10 mV
Gain adjustment range: NA
Gain accuracy: <±10 mV
Offset adjustment range: NA
D/A resolution: 12 bits (2.5 mV / LSB)
Sampling range: 250 s/sec
Output range: –5 V to +5 V
Noise: 5 mVpp at .01 to 1 kHz
Output impedance: 100 Ohms ±5 %
Frequency range: DC to 60 Hz (–3Db)
Environmental Requirements
Temperature range: Operating:10 to 40 °C (50 to 104 °F)
Storage: –20 to 50 °C (–4 to 122 °F)
Humidity: Operating:20 to 90 %, non-condensing
Storage: 10 to 95 % (with packaging)
Atmospheric pressure:
Operating:525 to 795 mmHg (70 to 106 kPa)
Storage: 375 to 795 mmHg (50 to 106 kPa)