B-30 DELTA/DELTA XL/KAPPA VF8
B T
ECHNICAL DATA
tpO
2
/CO
2
Pod
FiO
2
Sensors
Physical Attributes
Size (H x W x D): 52 x 132 x 153.4 mm (2.05 x 5.2 x 6.04 in)
Weight: 454 g (1 lb)
Connections: Connector for tpO2/tpCO2 solid-state electrode
Connector for pod-monitor interface cable
Interface Cable: 1m
Degree of protection against ingress of
water:
per IEC 60601-2-27 clause 44.3
Electrical Specifications
Mode of operation: Continuous
Power requirements: DC powered directly from the monitor
Protection against electric shock: Type CF
Environmental Requirements
Temperature range: Operating:15 to 40 °C (55 to 104 °F)
Storage: –20 to 50 °C (–-4 to 122 °F)
Relative humidity: Operating:20 to 95 %, non-condensing
Storage: 10 to 95 % (with packaging)
Atmospheric pressure: Operating:550 to 775 mmHg (73 to 103 kPa)
Storage: 375 to 795 mmHg (50 to 106 kPa)
Physical Attributes
Size (H x D): 40 x 30 mm (1.59 x 1.2 in)
Weight: 35 g (<1.2 oz)
Connections: NeoMed Pod, interface cable
Mounting: 16 mm thread x 1 mm pitch
Sensor type: Galvanic fuel cell (partial pressure)
Useful life: Approximately 1 year
NOTE: This sensor contains lead and caustic material. Dispose of or recycle properly and in
accordance with local regulations.
Environmental Requirements
Temperature range: Operating:10 to 40 °C (50 to 104 °F)
Storage: –10 to 50 °C (14 to 122 °F)
Relative humidity: Operating:20 to 90 %, non-condensing
Storage: 10 to 95 % (with packaging)
Atm. pressure: 600 to 900 mmHg (80 to 120 kPa)