20 Technical manual | X-am
®
2800 / X-am
®
5800
Troubleshooting
5.2 Warnings
The sensor technology that can be used depends on the device type.
The first digit of the code represents the sensor channel:
x32 Error in the compensation chan-
nel.
Switch the gas detector off and
then on again; replace the sensor.
x33 Error in a connected channel. Switch the gas detector off and
then on again; replace the sensor.
x38 Condensation on sensor Acclimatise the gas detector.
Move the gas detector into a con-
densation-free environment.
x39 Sensor software error Carry out a firmware update; con-
tact DrägerService.
x36 Device temperature too high. Operate gas detector within per-
mitted temperature range.
x37 Device temperature too low. Operate gas detector within per-
mitted temperature range.
x42 Sensor current too low. Switch the gas detector off and
then on again; replace the sensor.
Code Cause Remedy
Code Sensor channel Code Sensor channel
2xx DrägerSensor CatEx SR 6xx EC4
3xx EC1 7xx IR
4xx EC2 8xx IR
5xx EC3 9xx PID
Code Cause Remedy
150 Ambient pressure too high. Operate gas detector within per-
mitted ambient pressure range.
152 Customer’s period of use counter
will expire soon.
Reset period of use counter with
Dräger CC-Vision.
156 Ambient pressure too low. Operate gas detector within per-
mitted ambient pressure range.
157 X-am pump battery pre-alarm Charge the battery as soon as
possible. The battery will continue
for at least another 20 minutes
after the first battery pre-alarm.
159 Calibration not possible. Menu
function cannot be performed as a
notice exists that prevents the
function (e.g. sensors warming
up).
Determine notice code using the
Info menu and disable if neces-
sary.
163 Period of use expired. Reset period of use counter with
Dräger CC-Vision.
164 Bluetooth function limited. Switch the gas detector off and on
again. Contact DrägerService.