Maintenance
5.3 Instructions
6 Maintenance
WARNING
Risk of explosion!
To reduce the risk of ignition of a flammable or explosive atmosphere, observe the
following:
Do not open the gas detector in explosion-hazard areas.
WARNING
Danger to health!
Test gas may damage health if inhaled.
Do not inhale the test gas. Observe the hazard warnings of the relevant Safety
Data Sheets and the instructions for use of the gas detector! Observe the
national regulations when defining calibration intervals.
The operator of the product must protect the configuration software for the gas
detector (e.g. Dräger CC-Vision) with a password against unauthorized access as
part of their user management. The password must be changed regularly, at least
once per month.
Contact DrägerService to replace the display and printed circuit board.
6.1 Maintenance intervals
For inspection and maintenance, see e.g.:
EN/IEC 60079-29-2 – Gas detectors - Selection, installation, use and
maintenance of detectors for flammable gases and oxygen
EN 45544-4 – Electrical apparatus used for the direct detection and direct
concentration measurement of toxic gases and vapours - Part 4: Guide for
selection, installation, use and maintenance
National regulations
Code Cause
30 Wait for fresh air.
90 No sensor selected for this feature.
Check Interval
Inspections and maintenance by
experts.
Every 12 months
Check signalling elements with the sig-
nal test
Automatically whenever the device is
started