Technical manual | X-am
®
2800 / X-am
®
5800 27
Maintenance
X-am 5800: An active H2 added signal will be automatically deactivated
temporarily for the duration of a manual calibration, a PC calibration or an automatic
bump test.
6.7.2 Performing a fresh air calibration
Observe the following instructions for fresh air calibration:
● To improve accuracy, a fresh air calibration must be carried out if a zero
deviation exists.
● For the fresh air calibration, the display on the DrägerSensor XXS O
2 and XXS
O2 PR is set to 20.9 Vol%.
The fresh air calibration is not supported by the CO2 channel of the infrared
sensor and XXS O3. A zero calibration of these sensors can be undertaken by
performing a single-gas calibration with N2 or using the Dräger CC-Vision PC
software. A suitable zero gas is required here, one that is free from carbon dioxide
and ozone (e.g. N2).
Prerequisites
– The fresh air calibration can only be carried out if at least one sensor supports
fresh air calibration.
– The fresh air must be free of measured or interfering gases.
– The gas detector is switched on and warm-up phases 1 and 2 are complete.
To carry out a fresh air calibration:
1. Switch on the gas detector.
2. Call up fresh air calibration (depending on configuration):
Via the Quick menu:
● Press 3x in measuring mode.
● Select and confirm fresh air calibration .
Via the menu:
a. Press for approx. 4 s in measuring mode.
b. Enter and confirm the password.
c. Select and confirm fresh air calibration .
All measurement channels involved in the fresh air calibration flash.
3. Press to start fresh air calibration manually.
All measurement channels involved in the fresh air calibration flash.
4. If necessary, press to override the stability check. In this case, a calibration
takes place immediately.
Dräger recommends using the automatic stability control (wait until the gas
detector has independently carried out the calibration).
The new measured value is displayed for control purposes.
The result is displayed as follows:
OK fresh air calibration successful.
fresh air calibration failed.
OK
OK