42 Technical manual | X-am
®
2800 / X-am
®
5800
Configuration
6.14 Cleaning
The gas detector does not require any special care.
If it is heavily soiled, wash off the gas detector with cold water, using a sponge if
necessary. Dry the gas detector with a cloth.
NOTICE
Damage to the gas detector!
Harsh cleaning utensils (e.g. brushes), detergents and solvents can destroy the
dust and water filters.
Only clean the gas detector with cold water and, where necessary, a sponge.
7 Configuration
After a change to the configuration (e.g. with the CC-Vision PC software), check
the basic functions of the gas detector (e.g. alarm elements).
7.1 Standard gas configuration for X-am 2800
Different settings can be selected to meet customer requirements on delivery. The
current setting can be checked and changed with the Dräger CC-Vision PC
software.
The Dräger CC-Vision PC software can be downloaded at
www.draeger.com/software.
For information on suitable cleaning agents and disinfectants and their
specifications, see document 9100081 at
www.draeger.com/IFU.
DrägerSensor
Measuring
range
Alarm A1
- threshold
acknowledgeable
latching
Alarm A2
- thresh-
old
acknowledgeable
latching
CatEx SR [%LEL]
0 to 100 20
-
40
-
XXS O
2 [Vol%]
1)
1) There are 4 alarm thresholds for O2 2 ↑ (increasing) and 2 ↓ (decreasing).
0 to 25
↓19.5 | ↑22.
5
- ↓19 | ↑23 -
XXS O
2 PR [Vol%]
1)
2)
0 to 30
↓19.5 | ↑22.
5
- ↓19 | ↑23 -
XXS CO LC [ppm]
0 to 2000 30 - 60 -
XXS H
2S LC [ppm]
0 to 200 5 - 10 -
XXS NO
2 [ppm]
0 to 50 5 - 10 -
XXS SO
2 [ppm]
0 to 100 0.5 - 1 -