Technical manual | X-am
®
2800 / X-am
®
5800 43
Configuration
7.2 Standard gas configuration for X-am 5800
Different settings can be selected to meet customer requirements on delivery. The
current setting can be checked and changed with the Dräger CC-Vision PC
software.
The Dräger CC-Vision PC software can be downloaded at
www.draeger.com/software.
2) Only available to order as an option in the modular variant 3703900.
DrägerSensor
Measuring
range
Alarm A1
- threshold
acknowledgeable
latching
Alarm A2
- threshold
acknowledgeable
latching
CatEx SR [%LEL]
0 to 100 20
-
40
-
IR Ex/CO
2:
IR Ex [%LEL]
IR CO2 [Vol%]
0 to 100
0 to 5
20
0.5
-
-
40
1.0
-
-
IR Ex [%LEL] 0 to 100 20 - 40 -
IR CO
2 [Vol%] 0 to 5 0.5 - 1.0 -
XXS O
2 [Vol%]
1)
0 to 25
↓19.5 | ↑22.
5
- ↓19 | ↑23 -
XXS O
2 PR [Vol%]
1)
0 to 30
↓19.5 | ↑22.
5
- ↓19 | ↑23 -
XXS O2 100 [Vol%] 0 to 100 ↓18.5 | ↑24 - ↓17.5 | ↑25
-
XXS O
2/CO-LC
[Vol%] / [ppm]
0 to 25 O2
0 to
2000 CO
↓19 | ↑23 O
2
30 CO
-
-
↓17 | ↑24 O
2
60 CO
-
XXS O2 / H2S-LC
[Vol%] / [ppm]
0 to 25 O2
0 to 100 H2S
19 O2
5 H2S
-
-
23 O2
10 H2S
-
XXS CO LC [ppm]
0 to 2,000 30 - 60 -
XXS CO HC [ppm] 0 to 10,000 600
-
1200
-
XXS CO H
2-CP [ppm] 0 to 2,000 30
-
60
-
XXS H
2 [ppm] 0 to 2,000 200
-
400
-
XXS H
2 HC [Vol.-%] 0 to 4 0.8
-
1.6
-
XXS H
2S LC [ppm]
0 to 200 5 - 10 -
XXS H
2S HC [ppm] 0 to 1,000 10
-
20
-
XXS H
2S-LC/CO-LC
[ppm]
0 to 100 H2S
0 to 2000
CO
5 H2S
30 CO
-
-
10 H2S
60 CO -
XXS NO [ppm] 0 to 200 25
-
50
-
XXS NO
2 [ppm]
0 to 50 5 - 10 -