44 Technical manual | X-am
®
2800 / X-am
®
5800
Configuration
7.3 Configuring the gas detector
When a configuration is transmitted to the gas detector with the CC-Vision PC
software, existing TWA and STEL evaluations are reset.
Before carrying out a firmware update with the assistance of the PC software
CC-Vision, make sure that the battery is least 50 % charged. The most recent PC
software CC-Vision contains information concerning whether a firmware update is
available.
7.3.1 Configuring the gas detector with the PC and reading the data
memory
7.3.1.1 Connect the gas detector to a PC.
Equipment:
DIRA USB cable (order number 8317409)
Calibration adapter (order number 8318752)
XXS NO
2 LC [ppm] 0 to 50 0.5
-
1.0
-
XXS SO
2 [ppm]
0 to 100 0.5 - 1 -
XXS PH
3 [ppm] 0 to 20 0.1
-
0.2
-
XXS PH
3 HC [ppm] 0 to 2,000 5
-
10
-
XXS HCN [ppm] 0 to 50 1.9
-
3.8
-
XXS HCN PC [ppm] 0 to 50 5
-
10
-
XXS NH
3 [ppm] 0 to 300 20
-
40
-
XXS CO
2 0 to 5 0.5
-
1
-
XXS Cl
2 [ppm] 0 to 20 0.5
-
1
-
XXS OV [ppm] 0 to 200 10
-
20
-
XXS OV-A [ppm] 0 to 200 10
-
20
-
XXS Odorant [ppm] 0 to 40 10
-
20
-
XXS Amine [ppm] 0 to 100 10
-
20
-
XXS COCl
2 [ppm] 0 to 10 0.1
-
0.2
-
XXS O
3 [ppm] 0 to 10 0.1
-
0.2
-
PID HC neo 0.3 ... 2,000
Isobutylene
50 - 100 -
1) There are 4 alarm thresholds for O2 2 ↑ (increasing) and 2 ↓ (decreasing).
DrägerSensor
Measuring
range
Alarm A1
- threshold
acknowledgeable
latching
Alarm A2
- threshold
acknowledgeable
latching