EasyManua.ls Logo

Dräger Narkomed 2C - Page 130

Dräger Narkomed 2C
274 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ADJUSTMENT AND CALIBRATION PROCEDURES (continued) NM2C
SENSOR
CORD
CONNECTOR
RECEPTACLE
ON INTERFACE
PANEL
SV20316
SENSOR HOUSING
SENSOR CAPSULE
SENSOR HOUSING
COVER
INSPIRATORY
VALVE DOME
INSPIRATORY
VALVE DOME
PLUG
OXYGEN MONITOR SERVICE SCREEN
CURRENT CELL A: 366
CURRENT CELL B: 371
ZERO CALIBRATION PROCEDURE:
- REMOVE O2 CELL FROM SENSOR HOUSING
- LET CURRENT CELL VALUES STABILIZE
- SELECT "STORE ZERO" KEY
PRESS TRIGGER TO ENTER CALIBRATION VALUES
- REINSTALL OXYGEN CELL IN SENSOR HOUSING
STORED ZERO CELL A: 240
STORED ZERO CELL B: 241
STORE ZERO
SELECTS
FUNCTION
ACTIVATES
FUNCTION
EXIT
PRESSURE
MONITOR
Figure 5-5: OXYGEN MONITOR SERVICE SCREEN
Rev. G
5-14
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS

Table of Contents

Related product manuals