NM2C PMS PROCEDURE
6.0 PMS PROCEDURE, NARKOMED 2C
The procedures in this section shall be performed in their entirety each time a
component is removed, replaced, calibrated, adjusted and during all scheduled Periodic
Manufacturers Service (PMS) visits. A PMS Checklist form, available from the Draeger
Medical, Inc. Technical Service Department, shall be completed by the TSR each time
a PMS is performed. Steps in the procedure marked with ( ) require a response at the
corresponding line on the checklist form.
Space is also provided on the PMS checklist form to record the results of a vapor
concentration test. Contact the Draeger Medical, Inc. Technical Service Department for
vapor concentration verification procedures.
NOTE: Verify the dates on test equipment calibration labels. DO NOT USE any test
equipmenthavinganexpiredcalibrationdate.Notify yoursupervisorimmediately ifany
equipment is found to be out of calibration.
In the space provided at the bottom of the PMS checklist form, record the Model and
ID number of all calibrated test equipment used. Also record the calibration due dates.
Examples are: multimeter, digital pressure meter, Riken gas analyzer, safety analyzer,
volumeter, trace gas analyzer, simulators.
Test Equipment Required
:
-- Multi-Meter (Fluke or equivalent)
-- Electrical Safety Analyzer (Biotek 501 Pro or equivalent)
-- Regulator Test Pressure Gauge (dedicated for O
2
), P/N S000063 or equivalent
-- Regulator Test Pressure Gauge (for N
2
O and other gases), P/N S000063 or equivalent
-- Fresh Gas Outlet Volume Test Device, P/N S010158 or equivalent
-- Fresh Gas Leak Test Device, P/N 4113119 or equivalent
-- Adapter Assembly, Test Terminal, P/N 4104389 or equivalent: two are required
-- Flowmeter Test Stand (Capnomed), P/N S000081 or equivalent
-- Breathing System Leak Test Device, P/N S010159 or equivalent
-- Baromed Pressure Test Fixture or equivalent
-- Test Minute Volume Meter, P/N 2212300 or equivalent
-- Digital Pressure Manometer (SenSym PDM 200CD or Equivalent)
-- Riken Gas Indicator, Model 18H, or 1802D or equivalent
-- Stop Watch
-- Service Test Lung, P/N 4115128
Materials Required
:
-- Dow Corning High Vacuum Grease, P/N S4105908
-- Spiromed Sensor Lubrication Kit, P/N 2218180
-- Breathing Bag, 3 liter, P/N 9995330 or equivalent
-- Tube, Corrugated, 22 mm x 12 in. long, P/N 9995112
Test equipment illustrations are shown on following pages.
Rev. Q
6-1
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS