NM2C PMS PROCEDURE (continued)
( ) 6.3 BATTERY CIRCUIT
6.3.1 Is "ON" LED lighted? __ (Y)
6.3.2 With the System Power switch ON, unplug the AC power cord.
6.3.3 Is yellow "AC PWR FAIL" LED lighted as long as the power cord is
unplugged? __ (Y)
6.3.4 Does the "AC PWR FAIL" message appear in the Advisory display? __
(Y)
6.3.5 Press and hold the "BATTERY TEST" button.
6.3.6 Is green Battery Test LED lighted as long as "BATTERY TEST" button
is depressed? __ (Y)
6.3.7 Release the "BATTERY TEST" button.
6.3.8 Restore AC power to the machine.
6.3.9 Does the "AC PWR FAIL" message disappear and the LED extinguish?
( ) 6.4 CONFIGURATION
6.4.1 Press the Datagrip trigger, select SET and enter the SYSTEM
CONFIGURE screen.
6.4.2 The remote display shall display the Configuration Screen.
6.4.3 Verify the correct Time and Date.
6.4.4 Adjust the Volume to the highest setting.
Rev. K
6-5
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS