PMS PROCEDURE (continued) NM2C
6.5 SERVICE DATA
6.5.1 On the display key panel, press and hold the Speaker Disable and LOG
DATA or VENT ALARMS OFF keys (simultaneously), and then press
the Datagrip trigger (while still holding the keys).
6.5.2 The remote display shall display the Main Service Screen.
( ) 6.5.3 Record the Last Service Date on the PMS form.
( ) 6.5.4 Record the Hours Run Since Last Service on the PMS form.
( ) 6.5.5 Record the Total Hours Run on the PMS form.
6.5.6 Select and enter the Service Log.
6.5.7 Verify any pertinent information from the Service Log. Contact the
Draeger Medical, Inc. Technical Service Department if necessary.
6.5.8 Set the cursor to EXIT and press the trigger to return to the Main
Service Screen.
6.5.9 Select the PMS Service Code and press the Datagrip trigger.
6.5.10 Select and enter your Technical Service Rep. I.D. number.
( ) 6.5.11 Set the cursor to RESET and press the trigger. This resets the last
service date to the current date and resets the hours run since last
service to zero.
6.5.12 Set the cursor to PMS CRITERIA and press the trigger.
( ) 6.5.13 Select and enter the month of the next service due date. The internal
clock of the machine limits the amount of date advance to a maximum
of six months from the current service date.
Rev. K
6-6
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS