NM2C PMS PROCEDURE (continued)
( ) 6.10 VAPOR EXCLUSION SYSTEM
6.10.1 Set all vapors to zero (0).
6.10.2 Adjust the handwheel on the left vapor (viewed from the front of the
machine) to any concentration above zero (0).
6.10.3 Using the datagrip, select AGT.
6.10.4 Verify that the correct abbreviation for the agent selected appears on
the display.
6.10.5 Is it possible to adjust the center vapor? ___ (N)
6.10.6 Is it possible to adjust the right (viewed from the front of the machine)
vapor? ___(N)
6.10.7 Adjust the handwheel on the left vapor to zero (0).
6.10.8 Adjust the handwheel on the center vapor to any concentration above
zero 0).
6.10.9 Verify that the correct abbreviation for the agent selected appears on
the display.
6.10.10 Is it possible to adjust the left vapor? ___ (N)
6.10.11 Is it possible to adjust the right vapor? ___ (N).
6.10.12 Adjust the handwheel on the center vapor to zero (0).
6.10.13 Adjust the handwheel on the right vapor to any concentration above
zero (0).
6.10.14 Verify that the correct abbreviation for the agent selected appears on
the display.
6.10.15 Is it possible to adjust the left vapor? ___ (N).
6.10.16 Is it possible to adjust the center vapor? ___ (N).
6.10.17 Adjust the handwheel on the right vapor to zero (0).