NM2C PMS PROCEDURE (continued)
6.14.2 Air O.F.P.D. - if applicable
* 6.14.2.1 Set the gas selector switch to ALL GASES.
6.14.2.2 Open the air cylinder valve - if applicable.
6.14.2.3 Connect the air pipeline hose - if applicable.
6.14.2.4 Open one (1) oxygen cylinder valve.
6.14.2.5 Set the air flow to 4 l/min; set the oxygen flow to 4 l/min.
6.14.2.6 Close the oxygen cylinder valve.
( ) 6.14.2.7 Does the flow of air cease when the oxygen pressure is depleted?
___ (Y)
6.14.2.8 Close the air flow control valve.
6.14.2.9 Disconnect the air pipeline hose - if applicable.
6.14.3 Oxygen-Helium O.F.P.D. - if applicable
* 6.14.3.1 Set the gas selector switch to ALL GASES.
6.14.3.2 Open the oxygen-helium cylinder valve.
6.14.3.3 Open one (1) oxygen cylinder valve.
6.14.3.4 Set the oxygen-helium flow to 4 l/min; set the oxygen flow to 4
l/min.
6.14.3.5 Close the oxygen cylinder valve.
( ) 6.14.3.6 Does the flow of oxygen-helium cease when the oxygen pressure is
depleted? ___ (Y)
6.14.3.7 Close the oxygen-helium flow control valve.
* These items do not apply to machines with later design (P/N 4113329-001) ORC.