NM2C PMS PROCEDURE (continued)
6.16 O
2
MED
6.16.1 Disconnect the oxygen sensor cable from the Oxygen Sensor interface.
6.16.2 The following message shall appear on the remote display: O2 SENS
DISC.
6.16.3 Reconnect the O
2
Med sensor.
6.16.4 The following message shall appear on the remote display: CAL O2
SENSOR.
6.16.5 Press the Datagrip trigger, select SET UP and perform an O
2
calibration.
NOTE: Make sure that the sensor has been stabilized in ambient air for several
minutes.
( ) 6.16.6 After calibration is completed, what is the oxygen concentration? ___ %
(21)
6.16.7 Select TEMPLATE and press the trigger, select GET TEMPLATE and
enter the factory defaults.
6.16.8 The warning INSP O2 LOW shall appear on the remote display and the
warning heading shall be flashing. There shall be a continuous audible
alarm.
6.16.9 What is the low oxygen alarm default? ___ % (30)
6.16.10 Exit the Default screen and re-enter the SET UP screen.
6.16.11 Select the OXYGEN LOW alarm limit. Does a box appear around the
low alarm limit? ___ (Y)
6.16.12 Verify that the low alarm limit has a range from 18 to 99%.
6.16.13 Place the oxygen sensor into the inspiratory valve dome adapter, set the
Man/Auto selector BAG, close the APL valve. Attach a 12-inch hose to
the inspiratory valve and occlude the bag mount.
6.16.14 Set the oxygen flow to 4 l/min.
6.16.15 Set the low limit to 18, and verify the INSP O2 LOW message has
cleared.
Rev. K
6-37
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS