NM2C PMS PROCEDURE (continued)
6.20 BAROMED
6.20.1 Disconnect the breathing pressure sensor line from the absorber.
6.20.2 Connect a test pressure gauge and syringe to the breathing pressure
sensor line.
6.20.3 Enter the SET UP screen, and select the THRESHOLD PRES alarm
limit. Does a box shall appear around the Threshold Alarm Limit? ___
(Y).
6.20.4 What is the threshold alarm default? ___ cm H
2
O (12)
6.20.5 Verify that the threshold alarm limit has a range from 5 to 30 cm H
2
O.
6.20.6 Adjust the threshold to 10 cm H2O.
6.20.7 Select the PRESSURE HIGH alarm limit. Does a box shall appear
around the High Pressure Alarm Limit? ___ (Y)
6.20.8 What is the high alarm limit default? ___ cm H
2
O (50)
6.20.9 Verify that the high alarm limit has a range from 30 to 120 cm H
2
O.
6.20.10 Set the high alarm limit to 65 cm H
2
O, and exit from the set up menu.
6.20.11 Increase the pressure to 25 cm H
2
O, then decrease the pressure to 20
cm H
2
O.
6.20.12 Does the THRESHOLD LOW message appear in the Advisory column?
___ (Y)
6.20.13 Set the Man/Auto valve to AUTO, and turn the ventilator ON.
6.20.14 Bleed the pressure and start a stopwatch.
( ) 6.20.15 What is the time when APNEA-PRESSURE appears in the Caution
column? ___ sec (13-17)
( ) 6.20.16 What is the time whentheAPNEA-PRESSURE appears in theWarning
column? ___ sec (26-34)
6.20.17 After the APNEA-PRESSURE alarm is displayed in the Warning
column on remote display, slowly increase the test pressure.
( ) 6.20.18 At what pressure does the APNEA-PRESSURE alarm deactivate? ___
cm H
2
O (7-13)