TROUBLESHOOTING GUIDE (continued) NM2C
GUIDE 1: Loss of O
2
Monitor
N
Y
N
Y
N
Y
Y
N
Y
N
N
Y
START
UNIT IS
FUNCTIONAL
DOES UNIT
PASS PMS?
CONTACT NAD
SERVICE DEPT.
REINSTALL ORIGINAL
PROCESSOR ASSEMBLY
PERFORM A
COMPLETE PMS
ON UNIT
IS O2
MONITOR WORKING
PROPERLY?
REPLACE
PROCESSOR ASSEMBLY
AS OUTLINED IN
SECTION 4
IS O2
MONITOR WORKING
PROPERLY?
CHECK INTERNAL
CABLE FROM
INTERFACE PANEL
TO J1 ON
PROCESSOR BOARD
INSTALL SENSOR
ELEMENT CORRECTLY
CONNECT SENSOR
CORD TO
INTERFACE PANEL
IS SENSOR
ELEMENT INSTALLED
CORRECTLY?
IS
SENSOR CORD
CONNECTED TO
INTERFACE
PANEL?
IS O2
MONITOR WORKING
PROPERLY?
3-4
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS