SOFTWARE UPDATE PROCEDURE (continued) NM2C
7.6.1 Serial Port Configuration
The screen shown in Figure 7-4 indicates the current configuration parameters
for each of the Narkomed 2C serial ports, and also allows changing of these
settings.
-- At the SST Main Menu, set the cursor to SERIAL PORT
CONFIGURATION and press ENTER to display the Serial Port
Configuration screen.
MACHINE ID: NM2C
SERIAL NUMBER: 00011066
PT A
PT B
SV50102
SOFTWARE VERSION: 1.20
SYSTEM FEATURES: 0003
PROTOCOL
VITALINK
OR/LINK
BAUD
9600
9600
PARITY
NONE
EVEN
STOP
1
2
DATA
8
8
F10 PREVIOUS MENU
Figure 7-4: SERIAL PORT CONFIGURATION SCREEN
-- Move the flashing cursor in the PROTOCOL column to the line
corresponding to the port for which you want to change the settings.
Press ENTER, then use the up or down arrow key to make the selection, and
press ENTER again. The cursor will start flashing.
-- Move the flashing cursor to the BAUD column and press ENTER. Use the
up or down arrow key to select the desired baud rate, and press ENTER
again. The cursor will start flashing.
-- Move the flashing cursor to the PARITY column and press ENTER. Use
the up or down arrow key to select the desired parity, and press ENTER
again. The cursor will start flashing.