Instructions for Use Oxylog 2000 plus SW 1.n 107
Technical Data
Measured value display
Airway pressure measurement
Range 0 to 99 mbar
Resolution 1 mbar
Accuracy ±2 mbar
Maximum airway pressure PIP
Positive end expiratory pressure PEEP
Mean airway pressure Pmean
Plateau pressure Pplat
Flow measurement
Minute volume MVe
Range 0 to 99 L/min, BTPS
Resolution 0.1 L/min
Accuracy ±15% of measured value, or ±1 L/min, whichever is
greater.
Tidal volume VTe
Range 0 to 5000 mL, BTPS
Resolution 1 mL
Accuracy ±15% of measured value, or ±25 mL, whichever is
greater.
Frequency measurement
Range 0 to 99 /min
Resolution 1 /min
Accuracy ±1 /min
Pressure Bar graph
Airway pressure Paw –7 to 60 mbar