Instructions for Use Oxylog 2000 plus SW 1.n 35
Assembly
Supply from an O2 cylinder
1 Use a full O2 cylinder.
2 Connect the pressure reducer (270 to 600 kPa
delivery pressure, 500 kPa nominal pressure)
to the O
2 cylinder.
3 Connect the O2 medical gas hose (A) to the
Oxylog 2000 plus.
4 Connect the O
2 medical gas hose to the pres-
sure reducer (B).
5 Rotate the cylinder valve (C) slowly and open
fully.
Supply from a piped medical gas system
1 Connect the O2 medical gas hose (A) to the
Oxylog 2000 plus.
2 Connect the gas hose (B) to the O2 terminal unit
until the supply of O2 is confirmed.
WARNING
Only use compressed gas cylinders and pres-
sure reducers, which comply with all applicable
regulations and have been approved.
WARNING
Only use a pressure reducer with a relief valve at
the outlet to limit the delivery pressure to a max-
imum of 1000 kPa in case of a malfunction, to
prevent damage to the ventilator!
018
O
2
A
B
C
CAUTION
Do not connect flow control valves or flowmeters
in the gas supply to Oxylog 2000 plus.
The ventilator could malfunction!
WARNING
Always check the O2 pressure of cylinder
before use, to prevent insufficient oxylog sup-
ply during use.
019
A
B