Operation
48
Instructions for Use Oxylog 2000 plus SW 1.n
VC-CMV / VC-AC
VC-CMV Volume Controlled - Controlled Mecha-
nical Ventilation.
Volume-controlled ventilation with fixed mandatory
minute volume MV, set with tidal volume VT and
respiratory rate RR.
Set the ventilation pattern with the controls below
the display:
Tidal volume VT.
Ventilation respiratory rate RR.
(minimum possible respiratory rate: 5 per min).
Maximum airway pressure Pmax.
–O
2 setting, O2 AirMix or 100% O2 FiO2.
The following can be set on the display:
Positive end expiratory pressure PEEP.
Ventilation time ratio I:E.
Plateau time Tplat %, in % of the inspiration
time.
When setting the ventilation respiratory rate RR,
tidal volume VT or ventilation time ratio I:E, the
associated values for inspiration time Ti
and
inspiration flow are automatically displayed in the
information window.
025
WARNING
Only use VC-CMV for patients who are not
spontaneously breathing.
Otherwise, the patient may be put at risk by not
receiving sufficient ventilation.
Use VC-AC for patients with partial
spontaneous breathing.
Paw
Flow
Tinsp Te
Pplat
Plateau time
Tplat
1
RR
Insp. Flow
t
t
Pmax
PEEP
screen 3