Operation
52
Instructions for Use Oxylog 2000 plus SW 1.n
SpnCPAP (optional PS)
Continuous Positive Airway Pressure
Spontaneous breathing can optionally be assisted
with PS and NIV.
Set the ventilation pattern with the controls below
the display:
Maximum airway pressure Pmax.
–O
2 setting FiO2.
Apnea ventilation
Apnea back-up ventilation is only applicable when
using the SpnCPAP mode. In the event of an
apnea, the ventilator will automatically activate
volume-controlled mandatory ventilation
(VC-CMV).
When an apnea occurs, the device simultaneously
issues an alarm signal and switches to volume con-
trolled ventilation with the parameters respiratory
rate RRapn, tidal volume VTapn, and the maxi-
mum airway pressure Pmax when the apnea time
Tapn has been reached. The ventilation time ratio
I:E is set to 1:1.5. The plateau time Tplat % is 0.
The patient can breathe spontaneously during
apnea ventilation. The mandatory frequency
RRapn remains constant.
Setting apnea ventilation
On the display:
1 Press the Settings key (A) until
page appears.
2 Set Tapn with the rotary knob (B) to a value
between 15 and 60 seconds.
WARNING
Only use SpnCPAP for patients with sufficient
spontaneous breathing.
Otherwise there is a risk of the patient
receiving insufficient ventilation.
029US
Paw
Flow
Apnea alarm time
Tapn
Start
apnea ventilation
with apnea alarm
t
t
1
RR
Pmax
030
Oxylog 2000 plus
BC
A
2/3