Configuration
76
Instructions for Use Oxylog 2000 plus SW 1.n
Set the controls accordingly for the test:
VT to 500 mL
RR to 20 /min
Pmax to 40 mbar
FiO
2 to O2 AirMix
These settings are displayed on the screen.
To test the buttons:
1 Briefly press the corresponding button.
The associated letter on the screen changes
from "B" to "X". If the button has an LED, it will
be illuminated by the device. If there are buttons
without LED, the yellow warning LED will light
up on the device.
2 Briefly press the O key.
The ventilator switches OFF if it is pressed for
longer than 3 seconds.
The function of the rotary knob is not included in the
test.
Test loudspeaker, buzzer, LEDs and display
To test the loudspeaker, buzzer, all LEDs and the
display:
1 Select the required test
2 Start the test. Each function is tested by the
device.
To test the screen display (Test display):
3 Turn the rotary knob; various test cards are dis-
played.
The selected test remains active until the rotary
knob is pressed again.
Display accu (battery) and supply data
The parameters of the replaceable battery and the
status of the external power supply are displayed.
Display (example):
1 Advance to the second page:
2 Select the line Page, confirm and turn rotary
knob.
Display (example):
fd