Instructions for Use Oxylog 2000 plus SW 1.n 81
Problem Solving
!! Check settings
time
The expiration time result-
ing from the settings for RR
and I:E or Ti is not possible.
Change RR or I:E or Ti.13
!!! Confirm device
OFF with rotary
knob
Key
O has been pressed
for 3 seconds.
To switch OFF: confirm.
To continue ventilation, press key
O again.
---
!!! Device failure Technical defect. Contact your local DrägerService
for additional support.
1
!! Flow measure-
ment inop
Measurement hoses for
flow measurement hoses
kinked, disconnected or
leaking.
Ensure flow measurement hoses
are connected correctly.
23
Flow sensor defective. Replace flow sensor.
Technical defect. Contact your local DrägerService
for additional support – restricted
operation is now possible.
!! Gas delivery fail-
ure
Technical defect. Contact your local DrägerService
for additional support – restricted
operation is now possible.
20
!! High respiratory
rate
Patient breathes at a high
spontaneous rate.
Check patient's condition, check
ventilation pattern, correct alarm
limit RRsp if necessary.
15
!! Int. battery
charging inop
Technical defect. Contact your local DrägerService
for additional support – restricted
operation is now possible.
17
!!! Int. battery dis-
charged
The operating time for oper-
ation with the internal bat-
tery has expired and an
external DC supply has not
been connected.
The ventilator must immediately
be reconnected to a mains supply,
an on-board DC supply or a fully
charged battery.
2
!! Int. battery in use Oxylog 2000 plus draws its
power from the internal bat-
tery due to the absence of
an external DC supply.
Press the Alarm Reset key to
confirm the alarm.
12
!! Key failed Technical defect. Contact your local DrägerService
for additional support – restricted
operation is now possible.
19
Alarm Cause Remedy Alarm
Rank