Cleaning, Disinfection and Sterilization
88
Instructions for Use Oxylog 2000 plus SW 1.n
Disassembly
Disassemble the reusable hose system
1 Disconnect the ventilation hose (A) from the gas
output.
2 Disconnect the flow measuring hoses (B) from
the nozzles.
3 Disconnect the medical gas hose (C) from the
Oxylog 2000 plus.
011
B
A
C
011
WARNING
When disconnecting the ventilation hose,
always grip the sleeve (A) and not the corruga-
tions (B)!
If this is not done, the corrugations or hose
may be torn from the sleeve.
A
B